RTOG 0232 Plan Scoring
ETV Scoring: Arm 1 / I125: 110 Gy |
||
No Variation |
Minor Variation |
Major Variation |
Minimum Dose to 90% of
ETV is at, or above 99 Gy, and |
Minimum Dose to 90% of ETV is at, or above 88 Gy, or |
Failure to achieve either No Variation or Minor
Variation |
Minimum Dose to 90% of
ETV is less than 143 Gy. |
Minimum
Dose to 90% of ETV is at, or above 143 Gy. |
ETV Scoring: Arm 2 / I125: 145 Gy |
||
No Variation |
Minor Variation |
Major Variation |
Minimum Dose to 90% of
ETV is at, or above 130.5 Gy, and |
Minimum Dose to 90% of
ETV is at, or above 116.5 Gy, and |
Failure to achieve either No Variation or Minor
Variation |
Minimum Dose to 90% of
ETV is less than 188.5 Gy. |
Minimum Dose to 90% of
ETV is at, or above 188.5 Gy. |
ETV Scoring: Arm 1 / Pd103: 100 Gy |
||
No Variation |
Minor Variation |
Major Variation |
Minimum Dose to 90% of
ETV is at, or above 90 Gy, and |
Minimum Dose to 90% of
ETV is at, or above 80 Gy, and |
Failure to achieve either No Variation or Minor
Variation |
Minimum Dose to 90% of
ETV is less than 130 Gy. |
Minimum Dose to 90% of
ETV is at, or above 130 Gy. |
ETV Scoring: Arm 2 / Pd103: 125 Gy |
||
No Variation |
Minor Variation |
Major Variation |
Minimum Dose to 90% of
ETV is at, or above 112.5 Gy, and |
Minimum Dose to 90% of
ETV is at, or above 100 Gy, and |
Failure to achieve either No Variation or Minor
Variation |
Minimum Dose to 90% of
ETV is less than 162.5 Gy. |
Minimum Dose to 90% of
ETV is at, or above 162.5 Gy. |
Organs at Risk (OAR)
Urethra:
The maximum dose to the urethra and volume of
urethra (in cm3) that received more than 200% of the prescription
dose [U200(cc)] should be reported as well
.
Rectum: The outer rectal wall will be
contoured behind every axial slice of the prostate. The maximum dose to the
rectum and the volume of the rectum (in cm3) that received more than
100% of the prescription dose [R100(cc)]
should be reported.